Abstract

Epidemiologic studies have shown an association between mouse allergen exposure and asthma morbidity among urban populations, but confirmatory challenge studies in community populations have not been performed. This study was designed to examine the clinical relevance of mouse sensitization using a nasal challenge model. Forty-nine urban adults with asthma underwent skin-prick testing (SPT) and intradermal testing (IDT) with mouse epithelia extract. A positive SPT was defined as a net wheal size ≥3 mm and a positive IDT was defined as a net wheal size ≥6 mm using a 1:100 dilution of extract (1:10 w/v was obtained from Greer Laboratories (Lenoir, NC) as a single lot [Mus m 1 concentration = 2130 ng/mL]). Mouse-specific IgE (m-IgE) was measured by ImmunoCAP (Phadia, Uppsala, Sweden). Nasal challenge was performed with increasing concentrations of mouse epithelia extract and symptoms were assessed by visual analog scale. A positive challenge was defined as a 20-mm increase in the scale. The age range of the 49 participants was 18-50 years; 41% were men and 86% were black. Fourteen participants were SPT(+) to mouse, 15 participants were SPT(-) but (IDT(+)), and 20 participants were negative on both SPT(-) and IDT(-) (SPT(-)/IDT(-)). Sixty-four percent of the SPT(+) group, 40% of the IDT(+) group, and 20% of the SPT(-)/IDT(-) group had a positive nasal challenge. Sixty-seven percent (10/15) of those who were either SPT(+) or m-IgE(+) had a positive nasal challenge. SPT or the combination of SPT plus m-IgE performed best in diagnosing mouse allergy. The great majority of mouse-sensitized urban adults with asthma appear to have clinically relevant sensitization. Urban adults with asthma should be evaluated for mouse sensitization using SPT or SPT plus m-IgE testing.

Abstract

Background: Curcumin has been shown to decrease allergic symptoms and biomarkers in some animal and human studies.

Objective: Our study aimed to determine if curcumin affects immediate skin-prick testing.

Methods: We enrolled 34 participants sensitized to select antigens. The participants were randomized to treatment with curcumin or placebo in a double-blind fashion. The participants underwent titrated skin-prick testing before and after 1 week of treatment and the pre- and posttreatment skin test wheals and flares were compared.

Results: Curcumin did not have a statistically significant effect on immediate skin-prick test wheal or flare size.

Conclusion: Although curcumin may attenuate allergic symptoms and biomarkers, it does not have a significant effect on immediate skin-prick test results and does not need to be discontinued before testing.

Abstract

Background: Percutaneous allergen skin testing remains an established benchmark for diagnosing atopic disease. The reliability of skin testing depends greatly on the performance of allergen extracts used, methods used, and the presence of antihistamine medications.

Objective: To determine the differential effect of cetirizine on 2 different concentrations of histamine control solution and 5 common allergens used for percutaneous skin testing.

Methods: Twelve individuals underwent skin testing with histamine (1 and 6 mg/mL), control diluent, and 5 common aeroallergens. Wheal and flare measurements were measured in a masked fashion by a single operator. Cetirizine was administered for 4 consecutive days to determine the effect on both histamine and allergen wheal and flare responses.

Results: A total of 384 skin tests were performed on 12 volunteers. Cetirizine began to suppress wheal and flare responses at 1 hour (P < .05), with maximum suppression at day 5 (P < .05). Wheal and flare responses returned to greater than 90% baseline within 4 days of not taking cetirizine. Suppression was more apparent with 1 vs 6 mg/mL of histamine (62% vs 33%). Four of the 12 individuals taking cetirizine had a positive skin test result using 6 mg/mL of histamine control when the 1-mg/mL histamine test result was negative. Importantly, twice as many individuals had false-negative allergen responses using 6 mg/mL of histamine vs the 1 mg/mL as a positive control, although this finding did not reach statistical significance.

Conclusion: The use of a 6-mg/mL histamine control for some percutaneous skin test devices may result in more false-negative allergen responses because of the inability to detect the presence of antihistamines.

In this issue of the journal, Tversky et al present findings from a study comparing the performance of 10 different skin prick test devices and found that the histamine wheal sizes varied among devices, with some devices being more likely to result in a histamine wheal less than 3 mm. 1 As histamine was used as a surrogate for allergen, the findings suggested that some devices may have a higher probability of false negative results when allergen extracts are applied. Because reproducible skin testing is a major challenge in the practice of allergy, the authors are to be commended for studying the performance and reproducibility of skin testing devices.

To evaluate the skin test devices, Tversky et al conducted a prospective study of 24 participants and compared the wheal and flare size of histamine and control diluent at 6 locations on the upper and lower arms of the participants. The locations where the devices were applied were selected randomly. Wheal and flare responses were measured by independent, blinded technicians, and pain was assessed using the Wong-Baker pain rating scale. The authors defined false positive results as those where the wheal to the control diluent was both greater than or equal to 3 mm and no less than 2 mm smaller than the histamine wheal. By this definition, no device produced any false positives. The authors defined false negatives as those where the histamine wheal response was less than 3 mm and/or less than 2 mm larger than the diluent control. Generally, false negatives were uncommon, with all but 2 devices having less than or equal to 4% false negatives when the manufacturer's recommended histamine concentration was used. These 2 devices, the Quintip and GREER Pick, each had a false negative histamine response rate of 12%. Although false negative results attributable to the choice of a skin test device have been reported previously, the rate of false negative results in this study is somewhat higher than that reported in a similarly designed study that evaluated the performance of skin prick test devices using histamine and diluent controls. 2 However, the previous study used histamine concentrations of 10 mg/mL, which may explain the difference in false negative rates.

We do not know whether these findings mean that some devices result in high false negative allergen results. On one hand, it could be inferred that devices with more false negative histamine tests could also have more false negative allergen tests, leading to missed diagnoses of allergy. If this is true, it could have serious consequences, as false negative skin tests could result in inappropriate immunotherapy prescriptions or false reassurance about the chance of reacting to a food, insect sting, or animal. On the other hand, because both the histamine wheal and the allergen extract wheals are considered together when interpreting skin tests, the primary result of using a device with a higher false negative rate may be both a small histamine response and attenuated responses to the allergen extracts to which the patient has IgE. In this scenario, the results with the allergen extracts would not be interpreted as negative tests because the histamine response would also be poor, and instead would be considered uninterpretable and require either repeating the skin test or serum-specific IgE testing. Determining which of these interpretations is correct requires more research, but the general low false-negative rate of skin tests when compared against measures of clinical allergy such as oral food challenge outcome3 suggests that the second scenario may be more likely.

The second consideration is whether the device used needs to be taken into account when interpreting the skin test result. In their study, Tversky et al found widely varying average histamine wheal sizes. Although, again, allergen extracts were not used, it may be reasonable to infer that the wheal size to allergen extracts could similarly vary by device. A broader question, that has yet to be answered, is how different devices perform in terms of identifying clinically relevant sensitization. In most cases, the absolute allergen wheal size may not matter, but for foods, more and more studies are reporting predictive values for clinically relevant sensitization based on the food allergen wheal size. 3 Whether, and how, the device used needs to be considered when applying predictive values for clinically relevant sensitization to skin test results should be further studied.

The authors also evaluated the pain associated with the 10 devices and found no differences between multiheaded and single-headed devices. The Wong-Baker pain rating scale ranges from 1 to 10, and pain scores associated with the devices ranged from 2.04 to 3.29, so were low and similar. Although there were statistically significant differences in the pain score among devices, it is not clear if differences of the magnitudes observed here are clinically meaningful. The practical implications are that multipronged devices should not be avoided out of concern about pain and that no single device stands out as clearly superior (or inferior) in terms of pain.

Probably the most important conclusion of the study is that the devices resulted in variable wheal sizes and false negative rates when histamine was used as the test reagent. Although the rate of false negative results reported in this study is higher than those reported in similarly designed studies, 2 the general observation that the choice of device introduces variability in the skin test results is supported by several other studies that have addressed this question. However, there are many sources of variability in skin prick testing, so it is important that we understand all (or at least most) of them, and the relative contribution of each. As the authors point out, some sources of variability, in addition to the device, include interoperator variability, the site on which the skin tests are placed, recent use of antihistamines or other medications that affect the histamine response, provider interpretation, and, perhaps most important, unstandardized allergen extracts. If the 1 or 2 factors that contribute most to the variability in allergy skin tests are identified and the variability is removed, this would be a big step forward in improving the reproducibility of allergy skin test results. At the moment, though, it is not clear which factors contribute the most to skin test variability, or how big of a contributor the device is to the variability. It is also important to note that variability in detecting sensitization is only 1 factor in the imperfect predictive ability of skin prick testing and other diagnostic tests for allergy. IgE sensitization is common in the absence of clinically relevant disease, and all allergy testing needs to be interpreted in the context of the patient's histoty. Nonetheless, reducing variability in skin testing is an important first step toward refining it as a diagnostic tool.

Until we resolve these broader issues related to allergy skin testing, allergists should monitor the skin test results in their practices and be alert to unexpected patterns. If, for example, more patients than expected have inadequate histamine responses, it is well worth determining if the device is contributing to this pattern, and if so, to change to a device that results in larger histamine wheals. It is also important to review the histamine concentration being used as false negative histamine rates were lowest when the 6 mg/mL concentration was used. Otherwise, allergists should continue to judiciously select allergens to be tested according to the clinical scenario and interpret tests carefully.

1. Tversky JR, ChelJadurai Y. McGready J. Hamilton RG. Pe1form:111ce and pain tolerability of cun-ent diagnostic allergy skin prick test devices. J Allergy Clin lmmunol Pract 2015;3:888-93.

2. Nelson HS, Lahr J, Buchmeier A, McCormick D. Evaluation of devices for skin prick testing. J Allergy C!in lmmunol 1998:101:153-6.

3. Peters RL, Allen JU. Dhannage SC. Tang MLK. Koplin JJ, Ponsonby AL, et al. Skin prick test responses and allergen-specific lgE levels as predictors of peanut. egg, and sesame allergy in infants. J Allergy Clin lmmunol 20 I 3: I 32:874-80.

Division of Pediatric Allergy and Immunology, Johns Hopkins School of Medicine, Baltimore, Md 
No funding was received for this work. 
Conflicts of interest: E. C. Matsui has received research support from TherrnoFisher. C. A. Keet bas received research support from the National Institutes of Health.

Abstract


Background: Allergy skin test reliability depends on the reagents and controls selected. Histamine is used at 1mg/ml and 6mg/ml concentration but few studies address the rationale for selecting one versus the other and how this may impact diagnostic accuracy.

Objective: To determine the rate of false negative allergen skin tests responses between UniTest PC (using the 1mg/mL histamine) and Quintip devices (using 6mg/mL) for 4 common aeroallergens.

Methods: Subjects aged 18-65 with symptoms of allergy to cat and/or ragweed received skin testing with 4 aeroallergens (dust mite mix, timothy grass, ragweed, cat), histamine and control diluent. Those individuals who tested positive to cat or ragweed with one skin prick test (SPT) device but not the other then proceeded to nasal allergen challenge (NAC). The primary outcomes were the aeroallergen false negative rates and sensitivities of the skin test devices followed by nasal allergen (NAC).

Results: Twenty-five individuals were recruited and underwent a total of 300 SPTs. SPT to allergens (ragweed, dust mite, cat, and timothy grass) resulted in a statistically significant difference in wheal size among the two skin testing devices (p value <0.0001, 0.0001, 0.0006, and 0.0053 respectively). Six NAC procedures were performed to cat/ragweed and 5 of 6 (83% were positive). The overall allergen sensitivity rate for UniTest and Quintip were 97% and 78% respectively with most false negatives due to the use of 6mg/ml histamine control reagent.

Conclusion: Our study shows that 6mg/ml concentration of histamine control reagent may contribute to a false interpretation of aeroallergen skin prick test results.